ION CHAMBER DETECTOR, MODEL 6035

The model 6035 is an ion chamber detector. It is designed to be operated with the 6000 series components for readout. The detector is a stand alone component needing only 12 volts to operate. The electrometer and high voltage power supply are included. The sensitive electrometer is housed in a sealed housing to eliminate atmospheric and electrical noise. The output is a series of pulses proportional to the radiation field. This detector is designed for pulsed and steady state radiation fields.

Specifications

Mechanical
Size Overall: 9 in W x 9 in L x 12 in H excluding mounting bracket.
Construction: Case is Aluminum. Electrometer housing is gasketed cast aluminum. Desiccant is provided in housing for humidity control.
External connectors: Multipin for power and signal.

Detector
Detector type: Ion Chamber
Useful range: 6035: 1 mR/h to 1000 mR/h, 6035A: 10 mR/h to 10 R/h, 6035B: 0.1 mR/h to 100 mR/h
Sensitivity: 6035: 600 cpm/mR/h, 6035A: 60 CPM/mR/h, 6035B: 1200 CPM/mR/h
Suggested Alarm setting range: 10 times lowest end of useful range
Energy response: 80 Kev to 1600 Kev within 30%

Environmental
Temperature range: -10 to 50 C
Humidity range: 0 to 99% non condensing
   
   
Top of Page
Specifications subject to change without notice.