Home > Products > HPI > 6030 Data Sheet

Specifications
Bottom of page

LOW LEVEL ION CHAMBER DETECTOR
MODEL 6030

The model 6030 is an ion chamber detector. It is designed to be operated with the 6000 series components for readout. The detector is a stand alone component needing only 12 volts to operate. The electrometer and high voltage power supply are included. The sensitive electrometer is housed in a sealed housing to eliminate atmospheric and electrical noise. The output is a series of pulses proportional to the radiation field. This detector is designed for pulsed and steady state radiation fields.

Specifications

Mechanical
Size Overall: 10 in W x 18 in L x 7 in D excluding mounting bracket.
Construction: Case is Aluminum. Electrometer housing is gasketed cast aluminum. Desiccant is provided in housing for humidity control.
External connectors: Multipin for power and signal.

Detector
Detector type: Ion Chamber (Model 1055)
Useful range: 6030: 0.01 mR/h to 100 mR/h, 6030A: 0.1 mR/h to 1000 mR/h
Suggested Alarm setting range: 6030: 0.1 mR/h to 100 mR/h, 6030A: 1 mR/h to 1000 mR/h
Energy response: 80 Kev to 1600 Kev within 30%

Environmental
Temperature range: -10 to 50 C
Humidity range: 0 to 99% non condensing
   
   
Top of Page
Specifications subject to change without notice.
Far West Technology, Inc. © 2009 Fax: +1.805.964.3162 Tel: +1.805.964.3615 Email: [email protected]